CDH6000 – CALIBRATOR FOR H6000 LEAK DETECTOR

The CDH6000 is a gas H2N2 (5% hydrogen) supply that checks the calibration of the H6000.

It can be used with one or several calibrated leaks.

CDH6000 : FEATURES

• The CDH6000 must be fitted with a calibrated leak to deliver a known quantity of gas H2 N2 (5% hydrogen) at a given pressure.
• If the pressure displayed on the gauge of CDH6000 does not exceed 1 bar, change the cartridge (maximum pressure: 4 bar)
• To purchase cartridges filled with gas H2 N2 (5% hydrogen) contact ATEQ.

Easy Quick air leak calibrator
Hydrogen tracer gas tester

Tips : When the operator has finished using the calibrator CDH6000 disconnect the calibrated leak and store in its case and avoid the excessive pollution of the test area.

CDH6000 : OPERATIONS

Step 1
Push the catridge extention in Step 1.

CARTRIDGE EXTENSION

Step 2
The catridge is removed after extracting it out.

CARTRIDGE EXTRACTED

Step 3
Insert the new catridge into the extendor and click it.

To change the cartridge, remove it by pressing the button on the connector. Then, insert a new cartridge by sliding it down until it “clicks” into place

Step 4
Step 4 locking

Install the calibrated leak on the input connector of the CDH6000

Step 5
Step 5 Pressure adjustment
Adjust the pressure by pressing on the left to increase or right to decrease buttons until there is 1 bar on the gauge.

 

(1 bar=characteristic pressure value of the calibrated leak)

 